Plant Lighting Luminaire Spectrum and PPF Measurement

IMS-2021 Integrating Sphere Comprehensive Test System · PPF Plant Lighting Spectrum Measurement

Plant Lighting Luminaire Spectrum and PPF Measurement

Plant Lighting Luminaire Spectrum and PPF Measurement

Photosynthetic Photon Flux (PPF) is the core metric for evaluating plant lighting luminaire output, while spectral distribution determines the effect of light quality on plant physiology. The IMS-2021 integrating sphere comprehensive test system quickly and accurately collects full-space luminous flux, making it the standard configuration for laboratory and production line testing. This solution is based on the IMS-2021 integrating sphere system, using BaSO4 integrating sphere with SPM-5000 high-precision spectroradiometer to establish an accurate measurement method for spectral radiance and PPF of plant lighting LED luminaires, ensuring traceable data and compliance with DLC horticultural lighting technical requirements and industry standards.

To meet both spectral distribution and total photon flux measurement needs, the IMS-2021 system recommends the spectroradiometer method (SPM-5000) over the broadband probe method to avoid measurement errors from LED spectral shift. The BaSO4 integrating sphere's high-reflectance wall ensures efficient flux collection. The integrating sphere-spectroradiometer method for plant lighting luminaire spectrum and PPF measurement is the most rapid and accurate absolute flux measurement approach. Strict adherence to self-absorption correction and standard lamp calibration provides reliable data for R&D, factory QC and DLC horticultural lighting technical requirements certification.

Key Features

  • IMS-2021 integrating sphere comprehensive test system: BaSO4 sphere + SPM-5000 spectroradiometer; 380-800nm; avoids broadband probe misjudgment
  • Accurate PPF photosynthetic photon flux measurement: µmol/s; meets DLC horticultural lighting technical requirements V3.0/4.0
  • Self-absorption correction: auxiliary source corrects sphere self-absorption and luminaire absorption
  • Band integration: PAR (400-700nm), blue, green, red, far-red (700-800nm)
  • Complete reports: SPD plot, key parameters, band integration; traceable; DLC horticultural lighting compliant

Core Application Scenarios

Plant Lighting R&D - IMS-2021 PPF Precise Measurement

IMS-2021 integrating sphere system with BaSO4 sphere and SPM-5000 spectrometer rapidly measures PPF, PPE, spectral distribution and band integration for plant lighting luminaires; supports light recipe optimization and product finalization; DLC horticultural lighting compliant.

DLC Horticultural Lighting Certification - Technical Requirements Compliance

DLC horticultural lighting technical requirements V3.0/4.0 specify PPF and PPE test conditions and tolerances. IMS-2021 system outputs complete reports with SPD, PPF, PPE, band integration; traceable data for DLC certification.

Production Line QC - BaSO4 Sphere Rapid Testing

BaSO4 sphere high reflectance ensures flux collection efficiency; SPM-5000 enables fast spectral measurement. IMS-2021 system suits plant lighting luminaire production line sampling/full inspection for PPF, PPE and key parameter QC.

Measurable parameters

PPF Photosynthetic Photon Flux (µmol/s)
PPE Photosynthetic Photon Efficacy (µmol/J)
SPD Spectral Power Distribution, SQD Spectral Quantum Distribution
PAR (400-700nm), blue/green/red and far-red (700-800nm) photon flux
Input power, power factor PF, peak/dominant wavelength
Chlorophyll A/B weighted flux, R/B irradiance ratio

Technical specifications

  • IMS-2021 integrating sphere system: BaSO4 sphere; diameter per luminaire size (2m/1.5m/0.5m); wall reflectance >95%;
  • SPM-5000 spectroradiometer: 380-800nm (far-red to 780-800nm), ≤2nm resolution; DLC horticultural lighting compliant;
  • Auxiliary source: stable halogen or NIST/CNAS traceable standard lamp for self-absorption correction;
  • Fibre and detector: cosine-corrected head or transmissive cosine integrator for uniform angular response;
  • Electrical meter: power meter Class 0.2 or better, harmonic analysis; synchronous V, I, P, PF for PPE calculation.

Meet the standard requirements

  • IES LM-79-19 Electrical and Photometric Measurements of Solid-state Lighting Products
  • DLC Horticultural Lighting Technical Requirements V3.0/4.0

Test parameters

Key report parameters

  • Spectral power distribution plot
  • PPF (µmol/s), PPE (µmol/J), input power (W), PF, peak/dominant wavelength
  • PAR band (400-700nm) total photon flux
  • Blue/green/red/far-red photon flux and ratio

Test conditions

  • Ambient temperature, warm-up time
  • Sphere diameter, self-absorption correction applied