Medical Endoscope Cold Light Source Testing
Professional measurement system solutions

Medical Endoscope Cold Light Source Testing
The medical endoscope cold light source test solution is developed per YY 1081-2011 Medical Endoscope Cold Light Source standard. It measures relative spectral power distribution P(λ), chromaticity coordinates, CCT, CRI Ra and Ri, peak wavelength, FWHM, luminous flux, radiant flux, red/green/blue flux ratios φer/φeg and φeb/φeg, infrared cutoff φe/φ(300–1700 nm) and all YY 1081-2011 specified parameters for cold light source photometric and colorimetric measurement at the reference window.

Key Features
- Designed per YY 1081-2011 Medical Endoscope Cold Light Source standard
- Photometric and colorimetric measurement at reference window
- Meets YY1081 requirements: Ra≥90, CCT 3000K–7000K, red/green/blue flux ratios
- Measures infrared cutoff, spectral characteristics, radiant performance
Measurable parameters
Relative spectral power distribution P(λ)
Chromaticity coordinates (x,y), (u,v); CCT Tc; CRI Ra and Ri (I=1–15)
Peak wavelength, FWHM, luminous flux, radiant flux, color ratio, dominant wavelength, color purity
Red/green/blue flux ratios φer/φeg, φeb/φeg; green φeg, red φer, blue φeb
Infrared cutoff φe/φ(300–1700 nm)
Meet the standard requirements
- GB/T 5702-2019 Method for evaluating light source color rendering
- GB/T 7922-2023 Method of measuring color of lighting sources
- YY 1081-2011 Medical endoscope cold light source