Medical Endoscope Cold Light Source Testing

Professional measurement system solutions

Medical Endoscope Cold Light Source Testing

Medical Endoscope Cold Light Source Testing

The medical endoscope cold light source test solution is developed per YY 1081-2011 Medical Endoscope Cold Light Source standard. It measures relative spectral power distribution P(λ), chromaticity coordinates, CCT, CRI Ra and Ri, peak wavelength, FWHM, luminous flux, radiant flux, red/green/blue flux ratios φer/φeg and φeb/φeg, infrared cutoff φe/φ(300–1700 nm) and all YY 1081-2011 specified parameters for cold light source photometric and colorimetric measurement at the reference window.

Medical Endoscope Cold Light Source Testing

Key Features

  • Designed per YY 1081-2011 Medical Endoscope Cold Light Source standard
  • Photometric and colorimetric measurement at reference window
  • Meets YY1081 requirements: Ra≥90, CCT 3000K–7000K, red/green/blue flux ratios
  • Measures infrared cutoff, spectral characteristics, radiant performance

Measurable parameters

Relative spectral power distribution P(λ)
Chromaticity coordinates (x,y), (u,v); CCT Tc; CRI Ra and Ri (I=1–15)
Peak wavelength, FWHM, luminous flux, radiant flux, color ratio, dominant wavelength, color purity
Red/green/blue flux ratios φer/φeg, φeb/φeg; green φeg, red φer, blue φeb
Infrared cutoff φe/φ(300–1700 nm)

Meet the standard requirements

  • GB/T 5702-2019 Method for evaluating light source color rendering
  • GB/T 7922-2023 Method of measuring color of lighting sources
  • YY 1081-2011 Medical endoscope cold light source